$270 monthly/$90 weekly
A one-time setup/calibration charge in addition to rental rate noted above may also apply. When this applies, you’ll be informed of the amount of the additional charge at time of order or, if requested, prior to placement of order.
- Calibrator comes with in-line adaptor with quick disconnect for the calibration of the HV-1's
- Flow measuring device display's in CFM, LPM, and M3/min
- Range of 2 - 14 CFM (56 - 400 LPM)
- Flow rates are corrected to standard temperature and barometric pressure
- Display of barometric pressure and temperature in English units
MEASUREMENT PRINCIPLE
Flow Sensor:
Venturi Tube
Standardization:
Correction for standard temperature and barometric pressure
Curve linearization:
Individually calibrated and software corrected
MEASUREMENT RANGES
Range:
2 - 14 CFM (56 - 400 LPM)
Temperature:
-4° F - +122°F (-20°C +50° C)
Barometric pressure:
30 24 In-Hg (760 609 mm-Hg) approx. Sea level to 6000 ft. elevation above sea level
MEASUREMENT ACCURACY
Air flow:
+/- 2% of full scale (Optional +/- 1% of full scale)
Temperature (Typical):
1°C over range -4° F - +122°F (-20°C +50° C)
Barometric pressure:
+/- 1% over measurement range
DISPLAY:
6 Digit LCD 0.4" high
GENERAL INFORMATION
|
Power requirements: | 100120 VAC;50/60Hz | |
Operating temperature: | -4° F - +122°F | (-20°C +50° C) | |
Storage temperature: | -20° F - +122°F | -29°C +70° C) | |
Dimension (L×W×H) | 9.75"×8.25"×10.75" | (248×210×273mm) | |
Weight: | 8 lbs. 6 oz. | (3.8 Kgs.) | |
Installation Category: | Pollution Degree 2 | ||
Enclosure Rating: | IPXO |