$186 monthly/$62 weekly
A one-time setup/calibration charge in addition to rental rate noted above may also apply. When this applies, you’ll be informed of the amount of the additional charge at time of order or, if requested, prior to placement of order.
-
Simultaneous Alpha & Beta Counting
-
Independent Alpha & Beta
- Readouts
- Alarms Setpoints
- Audible Signatures -
Automatic Background Subtraction & Updating
-
50 Hour Battery Backup
APPLICABLE DETECTORS:
gas proportional, dual phosphor scintillator (like Ludlum Models 43-93, 43-68)
HIGH VOLTAGE:
200-2500 VDC
ALPHA PARAMETERS:
Threshold: 1-180 mV
Meter Scale: 0-1000 cpm
BETA PARAMETERS:
Threshold: 1-4 mV
Window: adjustable 4-70 mV
Meter Scale: 100-100,000 cpm
CROSSTALK:
? 1% alpha counts in beta channel, ? 10% gross counts in alpha channel
LINEARITY:
± 10% of true value, typically ± 2% of true value when measured with an electronic pulser
RESPONSE TIME:
beta varies between 1-5 seconds, alpha varies between 5-15 seconds
COUNT TIME:
adjustable between 0-120 seconds in 8 second increments
METERS:
dual meters (one each for alpha and beta), 1 mA, 2 inch DC movement
CONTROLS:
-
Front Panel:
- ACK/Reset button
- Meter readout model (HV, Alarm, Ratemeter) -
Rear Panel: power on/off
STATUS INDICATORS:
Green LEDs: updating background, checking, check on
Red LEDs: incomplete, overload, low battery
INPUTS:
count switch signal input via phone jack TINI-JAX #41
DETECTOR CONNECTOR:
series type C
POWER:
117 VAC 60 Hz, and internal rechargable 6 V (sealed-acid) battery, battery life approximately 50 hours
SIZE:
19.1 x 22.9 x 15.3 cm (7.5 x 9 x 6 in.) (H x W x L)
WEIGHT:
2.3 kg (4.9 lb.)