- Simultaneous Alpha/Beta Counting
- Dual LCD Display's with backlights
- Adjustable Count Time 0.1 - 30 minutes (or user defined value via PC software)
- Separate Speaker Tones for Alpha & Beta
- Maximum Sample Size of 2" Diameter X 0.4" thick
INDICATED USE:
Simultaneous Alpha/Beta sample counting
AUDIO:
Unimorph type speaker with volume control and dual tone (1 for each channel ) click-per-event audio
STATUS INDICATORS:
Backlit indicators for
QC-Daily QC check needed
OL-Scintillation detector is in overload condition
CPM/DPM-Counting in CPM or DPM mode
aAL/bAL-Count has exceeded alarm setpoint
SCALERS:
2 ea. 6 digit LCD displays with backlights with a range of 0 - 999999 counts (started by COUNT button)
SCALER LINEARITY:
Reading within ±2% of true value
COUNT TIMER:
Adjustable from 0.1 - 30 minutes (PC setting is user-defined via PC software)
HIGH VOLTAGE:
Adjustable from 200 - 2500 volts
THRESHOLD:
BETA - 4mV, ALPHA - 120 mV
BETA WINDOW:
50mV
DATA-OUTPUT:
9 pin RS-232 port
POWER:
85 - 265 VAC, 50-60 Hz single phase (less than 100 mA) with battery backup
CONSTRUCTION:
Beige polyurethane enamel paint and sub-surface printed front panel
TEMPERATURE RANGE:
-4°F (-20°C) to 122°F (50°C)>ELECTRONICS ONLY
SIZE:
8.5" (21.6 cm)H X 5" (12.7 cm)W X 9.3" (23.6 cm)D
WEIGHT:
Approximately 6 lbs (2.7 kg)
TUBE:
2" (5.1 cm) diameter magnetically shielded photomultiplier
WINDOW:
0.4 mg/cm² aluminized mylar
ACTIVE and OPEN AREA:
20.3 cm²
SAMPLE HOLDER:
Aluminum housing with sample tray capable of holding 1 or 2 diameter samples
EFFICIENCY(4pi geometry):
ALPHA: 37%-230Th; 39%-238U; 37%-239Pu
BETA: 8%-14C; 22%-99Tc; 29%-137Cs; 26%-90Sr/90Y
CROSS TALK:
Alpha to Beta: 10% or less
Beta to Alpha: 1% or less
BACKGROUND:
Alpha - 3 cpm or less Beta-Gamma - Typically 80 cpm or less (10 µR/hr field)
SOFTWARE:
PC based to perform setup and calibration routines including background subtract, crosstalk correction, cpm/dpm modes, daily QC check parameters, alarm levels, and automatic plateaus. All parameters are stored in the instrument in non-volatile memory. Logging of data is also available with software collecting and storing the following data: Sample Number, Date, Time, Alpha Count, Beta Count, Sample Type, & Comments